X-ray physics notes curriculum
Fundamentals of radiation
The X-ray machine
Production of X-rays
Interaction of radiation with matter
X-ray detection and image formation
Image quality
Radiation safety in X-ray imaging
Fluoroscopy (current module)
Mammography
In fluoroscopy, image brightness must remain constant despite variations in patient thickness or attenuation.
As the X-ray beam passes through regions of different density, the amount of radiation reaching the detector changes.
To maintain a consistent image appearance, the system automatically adjusts the exposure using a feedback mechanism known as Automatic Brightness Control (ABC) or, in modern systems, Automatic Dose Rate Control (ADRC).
Principle of Operation
Fluoroscopic systems incorporate a brightness sensor (photodiode or detector-based signal monitor) that continuously measures image luminance.
This signal is compared with a pre-set reference level corresponding to the desired image brightness. A feedback circuit adjusts the X-ray generator output to correct any deviation from the target value.
These adjustments occur in real time, typically once per frame, maintaining constant image brightness throughout the procedure.
Parameters Controlled
The system can maintain brightness by altering one or more exposure parameters depending on manufacturer design and the selected operating mode.
| Parameter | Effect on Image | Effect on Dose |
|---|---|---|
| kVp | Increases beam penetration | Slightly increases dose; reduces contrast |
| mA | Increases photon quantity | Directly increases dose |
| Pulse width | Increases exposure time per frame | Increases dose per frame |
| Pulse rate | Increases number of pulses per second | Increases cumulative dose |
Different modes prioritise parameters differently:
- Low-dose mode: favours increasing kVp before mA.
- High-contrast mode: favours mA adjustments for improved image contrast.
Region of Interest (ROI) Sensing
Brightness control is based on the signal detected from a defined region of interest (ROI) within the image field.
Most systems use a central weighted area, but digital fluoroscopy may allow user-defined ROIs. ROI sensing prevents bright or dark peripheral structures (e.g. contrast or metal) from disproportionately influencing exposure control.
Incorrect ROI selection can lead to either overexposure or underexposure of clinically relevant regions.
System Response
The ABC feedback loop responds within milliseconds in modern flat-panel systems, ensuring smooth transitions as patient thickness or field position changes. Older image intensifier systems may exhibit response lag, leading to transient over- or under-exposure when imaging rapidly changing anatomy.
System response characteristics are defined by control curves, which specify how exposure factors vary with patient attenuation.
Typically:
- kVp increases first to improve penetration.
- Once maximum kVp is reached, mA increases to maintain brightness.
Automatic Dose Rate Control (ADRC)
Modern fluoroscopic systems use Automatic Dose Rate Control (ADRC) instead of purely brightness-based control. ADRC regulates the dose rate incident on the detector to maintain a consistent signal-to-noise ratio (SNR), rather than simply maintaining optical brightness.
This provides a more stable image appearance and prevents unnecessary dose escalation when contrast agents or bright regions enter the field.
ADRC offers several selectable modes:
- Low dose / low frame rate: general positioning or guidance.
- Normal dose: standard diagnostic fluoroscopy.
- High-level / cine mode: brief high-quality acquisitions (e.g. angiography).
Relationship Between Brightness and Dose
Maintaining image brightness requires increasing tube output whenever patient attenuation increases. As a result, image brightness and patient dose are directly related.
| Scenario | System Response | Dose Effect |
|---|---|---|
| Beam moves from lungs to abdomen | ↑ kVp and mA | ↑ Dose |
| Use of magnification mode | ↓ input field → ↓ minification gain | ↑ Dose |
| Dense contrast or metallic object enters field | System compensates for apparent brightness loss | ↑ Dose |
Thus, while ABC maintains consistent image appearance, it can inadvertently increase dose if not monitored carefully. Modern systems mitigate this through dose-rate limits and real-time DAP displays.
Dose Management Features
ABC and ADRC systems often integrate with other dose optimisation features, including:
- Pulsed fluoroscopy: limits exposure duration per frame.
- Additional filtration: removes low-energy photons.
- Dose-rate limits: cap maximum tube current or kVp increases.
- Last-image hold: reduces unnecessary continuous exposures.
These measures maintain brightness stability while controlling cumulative dose.
Key Takeaways and Exam Tips:
- Automatic Brightness Control (ABC) maintains constant image brightness by adjusting kVp, mA, pulse width, or pulse rate.
- Automatic Dose Rate Control (ADRC) maintains consistent detector dose rate and SNR, improving efficiency in digital systems.
- ROI sensing confines brightness measurement to relevant anatomy.
- Increased brightness or magnification invariably raises patient dose.
- Common exam question: “Describe the operation of automatic brightness control in fluoroscopy and explain its effect on patient dose.”
Up Next
Next, we’ll move on to Dose Metrics and Dose-Saving Features, discussing how patient dose is measured, which metrics are displayed during fluoroscopy, and the techniques used to minimise dose while maintaining diagnostic image quality.