Mammography geometry and breast compression

Mammography employs a dedicated geometric design and controlled breast compression to optimise spatial resolution, reduce scatter, and minimise patient dose.

Precise alignment of the X-ray tube, compression paddle, and detector ensures consistent image quality and reproducible positioning across examinations.

System Geometry

Mammography systems use a dedicated, fixed geometry optimised for small field sizes and soft-tissue contrast.

ParameterTypical ValuePurpose / Effect
Source-to-image distance (SID)60–70 cmBalances magnification and geometric unsharpness
Anode angle6–12°Provides uniform coverage and small effective focal spot
Focal spot size0.3 mm (standard), 0.1 mm (magnification)Determines limiting spatial resolution
Field size~24 × 30 cmMatches average breast dimensions
Tube orientationCathode near chest wallCompensates for heel effect and tissue thickness
Line focus principle

The apparent focal spot is reduced by angling the anode.

This provides high spatial resolution while maintaining sufficient heat capacity in the actual focal track.

Heel Effect in Mammography

Due to the steep anode angle, the heel effect (variation in X-ray intensity across the field) is more pronounced in mammography.

Intensity decreases from the cathode side (chest wall) toward the anode side (nipple).

To ensure uniform image brightness:

  • The tube is positioned so that the cathode faces the chest wall, compensating for greater tissue thickness.
  • The anode side faces the thinner peripheral breast region.

This alignment balances beam intensity across the image receptor.

Compression: Purpose and Mechanism

The breast is compressed between a clear plastic paddle and the image receptor during every mammography exposure.

Compression is an essential design feature, not simply for positioning, but for image optimisation.

Benefits of compression include:

BenefitPhysical Principle / Outcome
Reduced tissue thicknessLess attenuation → lower dose and scatter
Reduced motion blurDue to immobilisation
Improved image contrastReduced scatter fraction → higher CNR
Uniform tissue thicknessReduces dynamic range requirements on detector
Better detail visibilityBrings anatomy closer to detector, reducing geometric unsharpness

Compression therefore directly improves image quality while simultaneously lowering radiation dose.

Typical Compression Parameters
ParameterTypical RangeNotes
Force applied≈100–200 NAdjusted automatically for comfort and breast size
Compressed thickness2–8 cm typicalDetermines kVp and filter selection via AEC
Exposure time0.5–1.5 sShorter due to reduced motion and tissue attenuation

Compression should be firm enough to achieve tissue uniformity and dose efficiency, but not so high as to cause excessive discomfort or risk tissue injury.

Magnification Mammography

Magnification views are used to evaluate microcalcifications or small lesions requiring fine detail.

Principle

Magnification is achieved by:

  • Increasing object-to-image distance (OID) by raising the breast above the detector.
  • Using a smaller focal spot (0.1 mm) to limit geometric unsharpness.
  • No anti-scatter grid (increased OID reduces scatter naturally).

Magnification improves spatial resolution at the expense of increased exposure and small loss of sharpness due to focal spot blur.

Scatter Reduction

In addition to compression, mammography employs focused anti-scatter grids to further improve contrast.

FeatureTypical Specification
Grid ratio4:1–5:1
Grid frequency30–50 lines/cm
Bucky factor≈2 (increases dose slightly)
Moving grid mechanismReduces grid-line artefacts

Some digital systems use slot scanning or software-based scatter correction, eliminating the need for a physical grid.

Key Points and Exam Tips:
  • Mammography systems use fixed geometry (SID ≈ 65 cm) with a small focal spot (0.1–0.3 mm).
  • The cathode side of the tube faces the chest wall to compensate for the heel effect.
  • Compression is essential for image quality and dose reduction.
  • Magnification views use small focal spots and increased OID for improved detail.
  • Scatter reduction is achieved through compression, grids, or digital correction.
  • Common exam question: “Explain the purpose of breast compression in mammography and describe its effects on image quality and radiation dose.”
Up next

Next, we’ll briefly discuss Image quality considerations in mammography. We’ve discussed image quality before, I just want to highlight key considerations in mammography.

Quick navigation

X-ray physics notes: